CVE-2020-15486 | Date: (C)2020-08-27 (M)2023-12-22 |
An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity.
CVSS Score and Metrics +CVSS Score and Metrics -CVSS V3 Severity: | CVSS V2 Severity: |
CVSS Score : 6.5 | CVSS Score : 3.3 |
Exploit Score: 2.8 | Exploit Score: 6.5 |
Impact Score: 3.6 | Impact Score: 2.9 |
|
CVSS V3 Metrics: | CVSS V2 Metrics: |
Attack Vector: ADJACENT_NETWORK | Access Vector: ADJACENT_NETWORK |
Attack Complexity: LOW | Access Complexity: LOW |
Privileges Required: NONE | Authentication: NONE |
User Interaction: NONE | Confidentiality: PARTIAL |
Scope: UNCHANGED | Integrity: NONE |
Confidentiality: HIGH | Availability: NONE |
Integrity: NONE | |
Availability: NONE | |
| |