CVE-2020-15486Date: (C)2020-08-27   (M)2023-12-22


An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity.

CVSS Score and Metrics +CVSS Score and Metrics -

CVSS V3 Severity:CVSS V2 Severity:
CVSS Score : 6.5CVSS Score : 3.3
Exploit Score: 2.8Exploit Score: 6.5
Impact Score: 3.6Impact Score: 2.9
 
CVSS V3 Metrics:CVSS V2 Metrics:
Attack Vector: ADJACENT_NETWORKAccess Vector: ADJACENT_NETWORK
Attack Complexity: LOWAccess Complexity: LOW
Privileges Required: NONEAuthentication: NONE
User Interaction: NONEConfidentiality: PARTIAL
Scope: UNCHANGEDIntegrity: NONE
Confidentiality: HIGHAvailability: NONE
Integrity: NONE 
Availability: NONE 
  
Reference:
https://payatu.com/advisory/lack-of-bluetooth-le-encryption-and-access-control-in-dr-trust-ecg-or-ekg-pen

CPE    1
cpe:/o:drtrust:electrocardiogram_pen_firmware:2.00.08
CWE    1
CWE-200

© SecPod Technologies