[Forgot Password]
Login  Register Subscribe

30481

 
 

423868

 
 

256148

 
 

909

 
 

199106

 
 

282

Paid content will be excluded from the download.


Download | Alert*
CVE
view JSON

CVE-2020-12035Date: (C)2020-06-30   (M)2023-12-22


Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The PrismaFlex device contains a hard-coded service password that provides access to biomedical information, device settings, calibration settings, and network configuration. This could allow an attacker to modify device settings and calibration.

CVSS Score and Metrics +CVSS Score and Metrics -

CVSS V3 Severity:CVSS V2 Severity:
CVSS Score : 4.9CVSS Score : 3.6
Exploit Score: 0.7Exploit Score: 3.9
Impact Score: 4.2Impact Score: 4.9
 
CVSS V3 Metrics:CVSS V2 Metrics:
Attack Vector: PHYSICALAccess Vector: LOCAL
Attack Complexity: LOWAccess Complexity: LOW
Privileges Required: LOWAuthentication: NONE
User Interaction: NONEConfidentiality: NONE
Scope: UNCHANGEDIntegrity: PARTIAL
Confidentiality: NONEAvailability: PARTIAL
Integrity: HIGH 
Availability: LOW 
  
Reference:
https://us-cert.cisa.gov/ics/advisories/icsma-20-170-02
https://www.us-cert.gov/ics/advisories/icsma-20-170-01

CPE    1
cpe:/h:baxter:prismaflex:-
CWE    1
CWE-798

© SecPod Technologies