CVE-2019-5024 | Date: (C)2019-04-15 (M)2023-12-22 |
A restricted environment escape vulnerability exists in the ���kiosk mode��� function of Capsule Technologies SmartLinx Neuron 2 medical information collection devices running versions 9.0.3 or lower. A specific series of keyboard inputs can escape the restricted environment, resulting in full administrator access to the underlying operating system. An attacker can connect to the device via USB port with a keyboard or other HID device to trigger this vulnerability.
CVSS Score and Metrics +CVSS Score and Metrics -CVSS V3 Severity: | CVSS V2 Severity: |
CVSS Score : 7.6 | CVSS Score : 7.2 |
Exploit Score: 0.9 | Exploit Score: 3.9 |
Impact Score: 6.0 | Impact Score: 10.0 |
|
CVSS V3 Metrics: | CVSS V2 Metrics: |
Attack Vector: PHYSICAL | Access Vector: LOCAL |
Attack Complexity: LOW | Access Complexity: LOW |
Privileges Required: NONE | Authentication: NONE |
User Interaction: NONE | Confidentiality: COMPLETE |
Scope: CHANGED | Integrity: COMPLETE |
Confidentiality: HIGH | Availability: COMPLETE |
Integrity: HIGH | |
Availability: HIGH | |
| |