[Forgot Password]
Login  Register Subscribe

30481

 
 

423868

 
 

256148

 
 

909

 
 

199106

 
 

282

Paid content will be excluded from the download.


Download | Alert*
CVE
view JSON

CVE-2018-8857Date: (C)2018-05-10   (M)2023-12-22


Philips Brilliance CT software (Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior) contains fixed credentials, such as a password or cryptographic key, which it uses for its own inbound authentication, outbound communication to external components, or encryption of internal data. An attacker could compromise these credentials and gain access to the system.

CVSS Score and Metrics +CVSS Score and Metrics -

CVSS V3 Severity:CVSS V2 Severity:
CVSS Score : 7.8CVSS Score : 7.2
Exploit Score: 1.8Exploit Score: 3.9
Impact Score: 5.9Impact Score: 10.0
 
CVSS V3 Metrics:CVSS V2 Metrics:
Attack Vector: LOCALAccess Vector: LOCAL
Attack Complexity: LOWAccess Complexity: LOW
Privileges Required: LOWAuthentication: NONE
User Interaction: NONEConfidentiality: COMPLETE
Scope: UNCHANGEDIntegrity: COMPLETE
Confidentiality: HIGHAvailability: COMPLETE
Integrity: HIGH 
Availability: HIGH 
  
Reference:
BID-104088
https://ics-cert.us-cert.gov/advisories/ICSMA-18-123-01
https://www.usa.philips.com/healthcare/about/customer-support/product-security

CWE    1
CWE-798

© SecPod Technologies