CVE-2018-14781 | Date: (C)2018-08-15 (M)2023-12-22 |
Medtronic MMT 508 MiniMed insulin pump, 522 / MMT - 722 Paradigm REAL-TIME, 523 / MMT - 723 Paradigm Revel, 523K / MMT - 723K Paradigm Revel, and 551 / MMT - 751 MiniMed 530G The models identified above, when paired with a remote controller and having the "easy bolus" and "remote bolus" options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.
CVSS Score and Metrics +CVSS Score and Metrics -CVSS V3 Severity: | CVSS V2 Severity: |
CVSS Score : 5.3 | CVSS Score : 2.9 |
Exploit Score: 1.6 | Exploit Score: 5.5 |
Impact Score: 3.6 | Impact Score: 2.9 |
|
CVSS V3 Metrics: | CVSS V2 Metrics: |
Attack Vector: ADJACENT_NETWORK | Access Vector: ADJACENT_NETWORK |
Attack Complexity: HIGH | Access Complexity: MEDIUM |
Privileges Required: NONE | Authentication: NONE |
User Interaction: NONE | Confidentiality: NONE |
Scope: UNCHANGED | Integrity: PARTIAL |
Confidentiality: NONE | Availability: NONE |
Integrity: HIGH | |
Availability: NONE | |
| |