CVE-2017-12714Date: (C)2018-04-26   (M)2023-12-22


Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted "RF wake-up" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. CVSS v3 base score: 5.3, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.

CVSS Score and Metrics +CVSS Score and Metrics -

CVSS V3 Severity:CVSS V2 Severity:
CVSS Score : 6.5CVSS Score : 6.1
Exploit Score: 2.8Exploit Score: 6.5
Impact Score: 3.6Impact Score: 6.9
 
CVSS V3 Metrics:CVSS V2 Metrics:
Attack Vector: ADJACENT_NETWORKAccess Vector: ADJACENT_NETWORK
Attack Complexity: LOWAccess Complexity: LOW
Privileges Required: NONEAuthentication: NONE
User Interaction: NONEConfidentiality: NONE
Scope: UNCHANGEDIntegrity: NONE
Confidentiality: NONEAvailability: COMPLETE
Integrity: NONE 
Availability: HIGH 
  
Reference:
BID-100523
https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01

CPE    1
cpe:/h:abbott:accent_mri:-

© SecPod Technologies