CVE-2015-3956 | Date: (C)2019-04-06 (M)2023-12-22 |
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
CVSS Score and Metrics +CVSS Score and Metrics -CVSS V3 Severity: | CVSS V2 Severity: |
CVSS Score : 9.8 | CVSS Score : 10.0 |
Exploit Score: 3.9 | Exploit Score: 10.0 |
Impact Score: 5.9 | Impact Score: 10.0 |
|
CVSS V3 Metrics: | CVSS V2 Metrics: |
Attack Vector: NETWORK | Access Vector: NETWORK |
Attack Complexity: LOW | Access Complexity: LOW |
Privileges Required: NONE | Authentication: NONE |
User Interaction: NONE | Confidentiality: COMPLETE |
Scope: UNCHANGED | Integrity: COMPLETE |
Confidentiality: HIGH | Availability: COMPLETE |
Integrity: HIGH | |
Availability: HIGH | |
| |