CVE-2014-5431Date: (C)2019-04-04   (M)2023-12-22


Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase. Baxter has released a new version of the SIGMA Spectrum Infusion System, version 8, which incorporates hardware and software changes.

CVSS Score and Metrics +CVSS Score and Metrics -

CVSS V3 Severity:CVSS V2 Severity:
CVSS Score : 6.8CVSS Score : 4.6
Exploit Score: 0.9Exploit Score: 3.9
Impact Score: 5.9Impact Score: 6.4
 
CVSS V3 Metrics:CVSS V2 Metrics:
Attack Vector: PHYSICALAccess Vector: LOCAL
Attack Complexity: LOWAccess Complexity: LOW
Privileges Required: NONEAuthentication: NONE
User Interaction: NONEConfidentiality: PARTIAL
Scope: UNCHANGEDIntegrity: PARTIAL
Confidentiality: HIGHAvailability: PARTIAL
Integrity: HIGH 
Availability: HIGH 
  
Reference:
https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01

CPE    1
cpe:/o:baxter:sigma_spectrum_infusion_system_firmware:6.05
CWE    1
CWE-798

© SecPod Technologies