[Forgot Password]
Login  Register Subscribe

30479

 
 

423868

 
 

250770

 
 

909

 
 

196157

 
 

282

Paid content will be excluded from the download.


Download | Alert*
CVE
view JSON

CVE-2014-5406Date: (C)2015-07-08   (M)2023-12-22


The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.

CVSS Score and Metrics +CVSS Score and Metrics -

CVSS V2 Severity:
CVSS Score : 9.3
Exploit Score: 8.6
Impact Score: 10.0
 
CVSS V2 Metrics:
Access Vector: NETWORK
Access Complexity: MEDIUM
Authentication: NONE
Confidentiality: COMPLETE
Integrity: COMPLETE
Availability: COMPLETE
  
Reference:
http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htm
https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01
https://xs-sniper.com/blog/2015/06/08/hospira-plum-a-infusion-pump-vulnerabilities/

CPE    2
cpe:/h:hospira:lifecare_pca5:-
cpe:/h:hospira:lifecare_pca3:-
CWE    1
CWE-345

© SecPod Technologies