[Forgot Password]
Login  Register Subscribe

30479

 
 

423868

 
 

248585

 
 

909

 
 

195621

 
 

282

Paid content will be excluded from the download.


Download | Alert*
CVE
view JSON

CVE-2019-18241Date: (C)2019-11-26   (M)2023-12-22


In Philips IntelliBridge EC40 and EC80, IntelliBridge EC40 Hub all versions, and IntelliBridge EC80 Hub all versions, the SSH server running on the affected products is configured to allow weak ciphers. This could enable an unauthorized attacker with access to the network to capture and replay the session and gain unauthorized access to the EC40/80 hub.

CVSS Score and Metrics +CVSS Score and Metrics -

CVSS V3 Severity:CVSS V2 Severity:
CVSS Score : 6.5CVSS Score : 3.3
Exploit Score: 2.8Exploit Score: 6.5
Impact Score: 3.6Impact Score: 2.9
 
CVSS V3 Metrics:CVSS V2 Metrics:
Attack Vector: ADJACENT_NETWORKAccess Vector: ADJACENT_NETWORK
Attack Complexity: LOWAccess Complexity: LOW
Privileges Required: NONEAuthentication: NONE
User Interaction: NONEConfidentiality: PARTIAL
Scope: UNCHANGEDIntegrity: NONE
Confidentiality: HIGHAvailability: NONE
Integrity: NONE 
Availability: NONE 
  
Reference:
https://www.us-cert.gov/ics/advisories/icsma-19-318-01

CWE    1
CWE-326

© SecPod Technologies