CVE-2018-4845Date: (C)2018-06-27   (M)2023-12-22


A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions < V3.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions >= V3.0 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =< V2.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions _with_ Siemens Healthineers Informatics products). Remote attackers with either local or remote credentialed access to the "Remote View" feature might be able to elevate their privileges, compromising confidentiality, integrity, and availability of the system. No special skills or user interaction are required to perform this attack. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue.

CVSS Score and Metrics +CVSS Score and Metrics -

CVSS V3 Severity:CVSS V2 Severity:
CVSS Score : 8.8CVSS Score : 6.5
Exploit Score: 2.8Exploit Score: 8.0
Impact Score: 5.9Impact Score: 6.4
 
CVSS V3 Metrics:CVSS V2 Metrics:
Attack Vector: NETWORKAccess Vector: NETWORK
Attack Complexity: LOWAccess Complexity: LOW
Privileges Required: LOWAuthentication: SINGLE
User Interaction: NONEConfidentiality: PARTIAL
Scope: UNCHANGEDIntegrity: PARTIAL
Confidentiality: HIGHAvailability: PARTIAL
Integrity: HIGH 
Availability: HIGH 
  
Reference:
https://cert-portal.siemens.com/productcert/pdf/ssa-755010.pdf

CWE    1
CWE-269