CVE-2018-18562Date: (C)2018-12-29   (M)2023-12-22


An issue was discovered in Roche Accu-Chek Inform II Base Unit / Base Unit Hub before 03.01.04 and CoaguChek / cobas h232 Handheld Base Unit before 03.01.04. Weak access credentials may enable attackers in the adjacent network to gain unauthorized service access via a service interface.

CVSS Score and Metrics +CVSS Score and Metrics -

CVSS V3 Severity:CVSS V2 Severity:
CVSS Score : 8.8CVSS Score : 3.3
Exploit Score: 2.8Exploit Score: 6.5
Impact Score: 5.9Impact Score: 2.9
 
CVSS V3 Metrics:CVSS V2 Metrics:
Attack Vector: ADJACENT_NETWORKAccess Vector: ADJACENT_NETWORK
Attack Complexity: LOWAccess Complexity: LOW
Privileges Required: NONEAuthentication: NONE
User Interaction: NONEConfidentiality: PARTIAL
Scope: UNCHANGEDIntegrity: NONE
Confidentiality: HIGHAvailability: NONE
Integrity: HIGH 
Availability: HIGH 
  
Reference:
BID-105843
https://ics-cert.us-cert.gov/advisories/ICSMA-18-310-01

CWE    1
CWE-521