[Forgot Password]
Login  Register Subscribe

30479

 
 

423868

 
 

249982

 
 

909

 
 

195748

 
 

282

Paid content will be excluded from the download.


Download | Alert*
CVE
view JSON

CVE-2017-9658Date: (C)2018-05-03   (M)2023-12-22


Certain 802.11 network management messages have been determined to invoke wireless access point blacklisting security defenses when not required, which can necessitate intervention by hospital staff to reset the device and reestablish a network connection to the Wi-Fi access point. During this state, the Philips IntelliVue MX40 Version B.06.18 can either connect to an alternative access point within signal range for association to a central monitoring station, or it can remain in local monitoring mode until the device is reset by hospital staff. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.

CVSS Score and Metrics +CVSS Score and Metrics -

CVSS V3 Severity:CVSS V2 Severity:
CVSS Score : 6.5CVSS Score : 6.1
Exploit Score: 2.8Exploit Score: 6.5
Impact Score: 3.6Impact Score: 6.9
 
CVSS V3 Metrics:CVSS V2 Metrics:
Attack Vector: ADJACENT_NETWORKAccess Vector: ADJACENT_NETWORK
Attack Complexity: LOWAccess Complexity: LOW
Privileges Required: NONEAuthentication: NONE
User Interaction: NONEConfidentiality: NONE
Scope: UNCHANGEDIntegrity: NONE
Confidentiality: NONEAvailability: COMPLETE
Integrity: NONE 
Availability: HIGH 
  
Reference:
BID-100813
https://ics-cert.us-cert.gov/advisories/ICSMA-17-255-01
https://www.usa.philips.com/healthcare/about/customer-support/product-security

CWE    1
CWE-755

© SecPod Technologies