CVE-2017-12722Date: (C)2018-02-15   (M)2023-12-22


An Out-of-bounds Read issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump reads memory out of bounds, causing the communications module to crash. Smiths Medical assesses that the crash of the communications module would not impact the operation of the therapeutic module.

CVSS Score and Metrics +CVSS Score and Metrics -

CVSS V3 Severity:CVSS V2 Severity:
CVSS Score : 5.3CVSS Score : 5.0
Exploit Score: 3.9Exploit Score: 10.0
Impact Score: 1.4Impact Score: 2.9
 
CVSS V3 Metrics:CVSS V2 Metrics:
Attack Vector: NETWORKAccess Vector: NETWORK
Attack Complexity: LOWAccess Complexity: LOW
Privileges Required: NONEAuthentication: NONE
User Interaction: NONEConfidentiality: NONE
Scope: UNCHANGEDIntegrity: NONE
Confidentiality: NONEAvailability: PARTIAL
Integrity: NONE 
Availability: LOW 
  
Reference:
BID-100665
BID-101252
https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A

CWE    1
CWE-125

© SecPod Technologies